Analytical methods for Irbesartan in bulk drug by Ultraviolet spectroscopy
Paras
Virani1, Kinjal Dungrani2
1Shree Dhanvantary Pharmacy College, Kim. Surat
2Globela Pharma Private Limited. Surat
*Corresponding Author E-mail: Parasvirani@gmail.com
ABSTRACT:
A simple, rapid and precise ultraviolet spectroscopy method was
developed and validated for quantitative analysis of irbesartan in bulk drug.
It is an angiotensin II receptor (AT1 subtype) antagonist. It is used for the
treatment of hypertension. It shows maximum absorbance at 246 nm. Beer
Lambert’s was obeyed in the concentration range of in the range of 5-30 μg/ml. The method was successfully validated in
accordance to USP pharmacopeia and ICH guideline for accuracy, precision, rang
and linearity. The linear regression analysis data for calibration plots showed
good linear relationship and obtain correlation factor is greater than 0.999
for irbesartan. The % Recovery/Accuracy was within the range between 98% and
102%. The percentage RSD for precision was found to be less than 2%. Thus
method was successfully applied for routine analysis of irbesartan in bulk
samples.
KEYWORDS: Accuracy, ICH guideline, Irbesartan,
Precision, Validation, ultraviolet spectroscopy USP pharmacopeia.
INTRODUCTION:
Irbesartan is in a group of drugs called angiotensin
II receptor antagonists. Irbesartan keeps blood vessels from narrowing, which
lowers blood pressure and improves blood flow. Irbesartan is used to treat high
blood pressure (hypertension). It may be used alone or in combination with
other antihypertensive agents. Irbesartan is a non-peptide compound, chemically
described as a 2-butyl-3-[p-(o-1H-tetrazol-5- ylphenyl)benzyl]-l ,3-diazaspiro [4.4]non-l -en-4-one. Its empirical
formula is C25H28N6O. Irbesartan is a white to
off white crystalline powder with a molecular weight of 428.5. Irbesartan is
slightly soluble in alcohol and methylene chloride
and practically insoluble in water [1-4]. The objective is to develop
ultraviolet spectroscopy
method and validate.
Validation is the process of generating experimental data that
provides evidence that the performance of an analytical method is adequate for
reliably assessing product quality [5-8]. The validation procedure has been
performed by UV spectroscopy. The method has been validated for linearity,
precision (system repeatability, method repeatability, and method
reproducibility), accuracy, range [9-14]. Ultraviolet spectroscopy refers to
absorption spectroscopy in the ultraviolet spectral region which ranges 200nm
to 400nm. Compounds which are colourless absorb
radiation in this region. In uv
spectroscopy, only valance electrons absorb energy, thereby the molecule
undergoes transition from ground state to exited state. This absorption is
characteristic and depends on the nature of electrons present. Molecules
containing π-electrons or nonbonding electrons (n-electrons) can absorb
the energy in the form of ultraviolet to excite these electrons to higher
anti-bonding molecular orbital. Medical options, several classes of drugs are
used for treatment of Hypertension. Actually now days cost of medicines are too
much high. There are two major ways to reduce the cost of medicine.
1) By route of synthesis
2) By analysis cost
Here an attempt has been made to reduce the cost of
medicine by reducing the analysis cost and develop such type of analytical
method in which, there is Minimum solvent consumption, reduced analysis time,
chemicals and reagents which are used in the method are cheap and easily
available. So the purpose of my research works to develop the analytical method
for anti-hypertensive pharmaceutical drugs by ultraviolet spectroscopy and
validate these methods with the guidance of United state Pharmacopeia (USP) and
International Conference on Harmonization (ICH).
MATERIALS AND METHODS:
Chemicals,
Reagents and instrumentation:
Pure samples of irbesartan were obtained as gift. LC
grade Methanol was purchased from Merck Company Mumbai. The analysis of the
drug was carried out on a Shimadzu UV-2450 Uv-Vis
spectrophotometer with single beam mono chromator.
Irbesartan shows maximum absorbance at 246 nm. Beer Lambert’s was obeyed in the
concentration range of in the range of 5-30 µg/ml. (Fig.-1)
Fig.3: Typical UV Scan spectra of
irbesartan 20ppm
Standard solution Preparation:
Stock A:- Accurately weigh and transfer 25mg of irbesartan working standard into a
100mL volumetric flask, add
about 70mL of
diluent and sonicate to dissolve it completely and make
volume up to the mark with the same.
5ppm solution: Take 1mL
of the above A solution into a 50mL volumetric flask and
dilute up to the mark with diluent.
10ppm solution: Take 2mL
of the above A solution into a
50mL volumetric flask and dilute up to the mark with diluent.
15ppm solution: Take 3mL
of the above A solution into a
50mL volumetric flask and dilute up to the mark with diluent.
20ppm solution: Take 4mL
of the above A solution into a
50mL volumetric flask and dilute up to the mark with diluent.
25ppm solution: Take 5mL
of the above A solution into a
50mL volumetric flask and dilute up to the mark with diluent.
30ppm solution: Take 6mL
of the above A solution into a
50mL volumetric flask and dilute up to the mark with diluent.
Assay of
irbesartan sample:- Accurately weigh and transfer equivalent to 25mg of irbesartan sample into a 100mL volumetric flask, add about 70mL
of diluent and sonicate to dissolve
it completely and
make volume up to the mark with
the same. Further pipette out 4mL of the
above stock solution into
a 50mL volumetric flask
and dilute up
to the mark
with diluent. Obtain solution concentration was 20μg/ml. absorbance
of irbesartan was
measured. Calculate % Irbesartan by following formulae.
Calculation
(A-I)
Irbesartan
(%)=------------X1000
SXC
Where,
A=Absorbance
of Irbesartan
in sample
I=Intercept
value obtained from linearity
curve
S= Slop value obtained
from
linearity curve
C= Concentration
of Irbesartan in
sample (mg/ml)
RESULTS
AND DISCUSSION:
Method validation:
The objective of
the method validation is
to demonstrate that
the method is suitable for its intended purpose as it is
stated in ICH guidelines. The method was
validated for linearity,
precision (repeatability and
intermediate precision), accuracy, specificity, stability and system
suitability. Linearity:- The linearity of an
analytical procedure is its ability to elicit test results that are directly,
or by a well-defined mathematical
transformation, proportional to the concentration of analyte
in sample within a given range. It should be established initially by visual
examination of a plot of signals as a function of analyte
concentration of content. If there appears to be a linear relationship, test
results should be established by appropriate statistical
methods (e.g., by calculation of a regression line by the method of least
squares). In some cases,
to obtain linearity between the response of an analyte and its concentration, the test data may
have to be subjected to a mathematical
transformation. Data from the regression line
itself may be helpful for providing mathematical estimates of the
degree of linearity. The
correlation coefficient, y-intercept, slope of the regression line, and
residual sum of squares should be submitted.
Six standard
solutions of Irbesartan
were prepared from three stocks in the range of 25% to 150
% of the nominal concentration
and taken absorbance
once. Linearity regression analysis demonstrated the acceptability of
the method for quantitative determination of
Irbesartan over the concentration
range of about 5ppm to 30ppm of the nominal concentration. Linearity graph was
shown in Fig.2 and slope, intercept correlation factor and Regression equation
were shown in Table.1
Fig.2: Linearity graph
Table1:LinearityData
|
Irbesartan Concentration (ppm) |
Irbesartan Area |
|
5.04ppm |
0.1699 |
|
10.01ppm |
0.3315 |
|
15.04ppm |
0.5014 |
|
20.17ppm |
0.6655 |
|
25.02ppm |
0.8387 |
|
30.08ppm |
0.9966 |
|
Slope |
33.1022 |
|
Intercept |
0.0019 |
|
Correlation factor |
0.9999 |
Precision:-
The precision of
an analytical method is the degree of agreement among individual test results when the method is applied repeatedly to multiple samplings of a
homogeneous sample. The precision of an
analytical
method is usually
expressed as the standard deviation or relative standard
deviation (coefficient of variation) of a series of measurements. Precision may be a
measure of either the degree of reproducibility or repeatability of the
analytical method under normal operating conditions. The ICH documents recommend that repeatability should be assessed using a minimum of nine determinations covering the specified range
for the procedure (i.e., three
concentrations and three replicates of
each concentration, or a minimum of six determinations at 100% of the test
concentration).
System Repeatability:-
Standard solution is prepared at three difference concentration
form different stock solution 15ppm, 20ppm and 25ppm were analysed
and taken absorbance on triplicate and RSD of absorbance at each level
were calculated. The percentage RSD of absorbance was less
than 1.0%. Result are presented in Table.2
Table2:System Repeatability data
|
System Repeatability |
||
|
SNo. |
Concentration
(ppm) |
Absorbance |
|
1 |
15.04 |
0.5032 |
|
2 |
0.5066 |
|
|
3 |
0.5029 |
|
|
|
RSD (%) |
0.4 |
|
1 |
20.17 |
0.6737 |
|
2 |
0.6756 |
|
|
3 |
0.6737 |
|
|
|
RSD (%) |
0.2 |
|
1 |
25.02 |
0.8387 |
|
2 |
0.8349 |
|
|
3 |
0.8276 |
|
|
|
RSD(%) |
0.7 |
Method repeatability:-
Six preparation of Irbesartan sample was analyzed from sample
preparation to final results by the same analyst and the percentage RSD of
obtained results was less than 2% and obtained result were within given range
100 ± 2. Result are presented in Table.2
System Reproducibility:-
Three Irbesartan sample are analysed by
this method in duplicate preparation and obtain result are in Table.4
Accuracy:-
It is defined as the closeness of test results obtained by the
method to the true value. It may often be expressed as percent recovery by
the assay of known, added
amounts of analyte. Accuracy is
a measure of the exactness of the analytical method. The ICH documents
recommend that accuracy be assessed using a minimum of nine determinations over a minimum of three
concentration levels, covering the
specified range (i.e., three
concentrations and three replicates of
each concentration). The three different concentrations of
Irbesartan standard solutions were determined
from three replicate analysis,
using the linear regression lines (linearity section). The deviations of the obtained results
(expressed as percentage accuracy) were calculated from the true values were presented
in Table.5 The average deviations from true value are less than 2.0 %.
Range:-
The range obtained from Linearity, Precision and Accuracy
is summarized -
ibesartan-5 ppm to 30ppm
(25% to 150%
of nominal sample concentration).
CONCLUSION
The method validation demonstrated that The Method is selective,
precise, linear, and accurate for over the required concentration ranges of 5
to 30 ppm (25 to150 % of Irbesartan nominal
sample concentration) for quantitative analysis of irbesartan by UV
spectroscopy. So this method should
be used in routine for quantitative analysis of
irbesartan.
REFERENCE
1.
The Merck Index, An
Encyclopaedia of Chemicals, Drugs and Biological,
Thirteenth Edition, Merck & Co.
Inc., 2001, 13:3453.
2.
P Palatini. The Journal of
Clinical Hypertension., 2005, 7,
96–101.
3.
USP 32:
"Validation of Compendial
Methods".
4.
Guidance for
Industry – Analytical Procedures and Method Validation (August-2000).
5.
W.D. Snyder, L.
Blumberg, in: P. Sandra, M.L.
Lee (Eds.), Proceedings of the 14th International Symposium on
Capillary Chromatography,
Baltimore, MD, May 1991, p. 28
6.
WHO, Guidelines
for Stability Testing
of Pharmaceutical Products Containing Well Established Drug Substances in Conventional Dosage Forms, in WHO Expert
Committee
7.
Willard Hobart.
H., Merritt L.L., Dean John. A., Instrumental Methods of Analysis, 7th edition, CBS Publishers, 580-610.
8.
International
Conference on
Harmonization of Technical Requirement for Registration of
Pharmaceuticals for Human
use, ICH harmonized tripartite
Guideline, Validation of Analytical procedures Text and methodology Q2 (R1), 2005.
9.
R. Jenke, “Chromatographic Method
Validation: A review of
Current Practices and
Procedures. I. General Concepts
and Guidelines”, J. Liq. Chrom. and
Rel. Technol., vol. 19 (1996), pp. 719-736.
10. Albero, I., Rodenas, V., Garcia,
S., and Sanchez, Pedreno,
C., “Determination of Irbesartan in the presence of Hydrochlorothiazide by
derivative Spectrophotometry”, J. Pharm. Biomed. Anal., 2004, 29 (1-2), pp.
229-305.
11. Vetuchi, C., Giannandrea, A. Carlucci, G., and Mazzeo, P.”Determination of hydrochlorothiazide and
Irbesartan in
Pharmaceuticals by fourth-order
UV derivative Spectrophotometry,” Farmaco, 2005, 60(8), pp.665-670.
12. Gagigal E, Gonzalez
L, Alonso RM,
Jimenez RM. Experimental
design methodologies to optimise the spectrofluorimetric
determination of Losartan and Valsartan in
human urine. Talanta 2001;54:1121-33.
13. Wang L, Asgharnejad M.
Second derivative UV spectrometric determination of simvastatin
in its tablet dosage form.
J Pharm Biomed
Anal 2000;21:1243-1248.
14. Surekha A, Jain NK. Difference spectrophotometric estimation of amlodipine besylate.
Indian drugs 2000;37(7):351-353.
Received on 17.09.2015 Accepted on 20.10.2015
© Asian Pharma
Press All Right Reserved
Asian
J. Pharm. Ana. 5(3): July- Sept. 2015; Page 147-150
DOI: 10.5958/2231-5675.2015.00024.1