Analytical methods for Irbesartan in bulk drug by Ultraviolet spectroscopy

 

Paras Virani1, Kinjal Dungrani2

1Shree Dhanvantary Pharmacy College, Kim. Surat

2Globela Pharma Private Limited. Surat

*Corresponding Author E-mail: Parasvirani@gmail.com

 

ABSTRACT:

A simple, rapid and precise ultraviolet spectroscopy method was developed and validated for quantitative analysis of irbesartan in bulk drug. It is an angiotensin II receptor (AT1 subtype) antagonist. It is used for the treatment of hypertension. It shows maximum absorbance at 246 nm. Beer Lambert’s was obeyed in the concentration range of in the range of 5-30 μg/ml. The method was successfully validated in accordance to USP pharmacopeia and ICH guideline for accuracy, precision, rang and linearity. The linear regression analysis data for calibration plots showed good linear relationship and obtain correlation factor is greater than 0.999 for irbesartan. The % Recovery/Accuracy was within the range between 98% and 102%. The percentage RSD for precision was found to be less than 2%. Thus method was successfully applied for routine analysis of irbesartan in bulk samples.

 

KEYWORDS: Accuracy, ICH guideline, Irbesartan, Precision, Validation, ultraviolet spectroscopy USP pharmacopeia.

 

 


INTRODUCTION:

Irbesartan is in a group of drugs called angiotensin II receptor antagonists. Irbesartan keeps blood vessels from narrowing, which lowers blood pressure and improves blood flow. Irbesartan is used to treat high blood pressure (hypertension). It may be used alone or in combination with other antihypertensive agents. Irbesartan is a non-peptide compound, chemically described as a 2-butyl-3-[p-(o-1H-tetrazol-5- ylphenyl)benzyl]-l ,3-diazaspiro [4.4]non-l -en-4-one. Its empirical formula is C25H28N6O. Irbesartan is a white to off white crystalline powder with a molecular weight of 428.5. Irbesartan is slightly soluble in alcohol and methylene chloride and practically insoluble in water [1-4]. The objective is to develop ultraviolet spectroscopy method and validate.

 

Validation is the process of generating experimental data that provides evidence that the performance of an analytical method is adequate for reliably assessing product quality [5-8]. The validation procedure has been performed by UV spectroscopy. The method has been validated for linearity, precision (system repeatability, method repeatability, and method reproducibility), accuracy, range [9-14]. Ultraviolet spectroscopy refers to absorption spectroscopy in the ultraviolet spectral region which ranges 200nm to 400nm. Compounds which are colourless absorb radiation in this region. In uv spectroscopy, only valance electrons absorb energy, thereby the molecule undergoes transition from ground state to exited state. This absorption is characteristic and depends on the nature of electrons present. Molecules containing π-electrons or nonbonding electrons (n-electrons) can absorb the energy in the form of ultraviolet to excite these electrons to higher anti-bonding molecular orbital. Medical options, several classes of drugs are used for treatment of Hypertension. Actually now days cost of medicines are too much high. There are two major ways to reduce the cost of medicine.

 

1) By route of synthesis

2) By analysis cost

 

Here an attempt has been made to reduce the cost of medicine by reducing the analysis cost and develop such type of analytical method in which, there is Minimum solvent consumption, reduced analysis time, chemicals and reagents which are used in the method are cheap and easily available. So the purpose of my research works to develop the analytical method for anti-hypertensive pharmaceutical drugs by ultraviolet spectroscopy and validate these methods with the guidance of United state Pharmacopeia (USP) and International Conference on Harmonization (ICH).

 

MATERIALS AND METHODS:

Chemicals, Reagents and instrumentation:

Pure samples of irbesartan were obtained as gift. LC grade Methanol was purchased from Merck Company Mumbai. The analysis of the drug was carried out on a Shimadzu UV-2450 Uv-Vis spectrophotometer with single beam mono chromator. Irbesartan shows maximum absorbance at 246 nm. Beer Lambert’s was obeyed in the concentration range of in the range of 5-30 µg/ml. (Fig.-1)

 

Fig.3: Typical UV Scan spectra of irbesartan 20ppm

 

Standard solution Preparation:

Stock A:- Accurately weigh  and transfer 25mg  of irbesartan working standard into a 100mL  volumetric flask,  add   about  70mL   of  diluent  and   sonicate  to dissolve it completely and   make   volume   up  to the mark with the same.

5ppm solution: Take  1mL  of  the above  A solution into a 50mL volumetric flask and dilute up to the mark with diluent.

10ppm solution: Take  2mL  of the above  A solution into a 50mL volumetric flask and dilute up to the mark with diluent.

15ppm solution: Take  3mL  of the above  A solution into a 50mL volumetric flask and dilute up to the mark with diluent.

20ppm solution: Take  4mL  of the above  A solution into a 50mL volumetric flask and dilute up to the mark with diluent.

25ppm solution: Take  5mL  of the above  A solution into a 50mL volumetric flask and dilute up to the mark with diluent.

30ppm solution: Take  6mL  of the above  A solution into a 50mL volumetric flask and dilute up to the mark with diluent.

 

Assay  of irbesartan sample:- Accurately  weigh  and transfer equivalent to 25mg  of irbesartan sample into a 100mL  volumetric flask, add  about 70mL  of diluent and   sonicate  to  dissolve  it  completely  and   make volume  up to the mark with the same.  Further pipette out 4mL  of  the above   stock solution  into  a  50mL volumetric  flask   and   dilute  up   to  the  mark  with diluent. Obtain solution concentration was 20μg/ml. absorbance of irbesartan was  measured. Calculate % Irbesartan by following formulae.

 

Calculation

                                   (A-I)

Irbesartan (%)=------------X1000

                                  SXC

Where,

A=Absorbance of Irbesartan in sample

I=Intercept value obtained from linearity curve

S= Slop value obtained from linearity curve

C= Concentration of Irbesartan in sample (mg/ml)

 

RESULTS AND DISCUSSION:

Method  validation: 

The   objective  of  the  method validation  is   to  demonstrate  that  the  method  is suitable for its intended purpose as it is stated in ICH guidelines.  The  method  was   validated  for linearity, precision (repeatability and  intermediate precision), accuracy, specificity, stability and system suitability. Linearity:- The linearity of an analytical procedure is its ability to elicit test results that are directly, or by a well-defined mathematical          transformation, proportional to the concentration of analyte in sample within a given range. It should be established initially by visual examination of a plot of signals  as a function of analyte concentration of content. If there appears to be   a   linear  relationship,   test   results  should  be established by appropriate statistical methods (e.g., by calculation of a regression line by the method of least squares).  In some  cases,  to obtain linearity between the response of an  analyte and  its concentration, the test data may have  to be subjected to a mathematical transformation.  Data from the  regression  line  itself may  be helpful  for providing mathematical estimates of the degree of linearity. The  correlation coefficient, y-intercept, slope  of the regression line,  and  residual sum of squares should be submitted.

 

Six standard  solutions of  Irbesartan were prepared from three stocks in the range of 25%  to 150  % of the nominal concentration  and   taken  absorbance  once. Linearity regression analysis demonstrated the acceptability of the method for quantitative determination  of  Irbesartan  over the concentration range of about 5ppm to 30ppm of the nominal concentration. Linearity graph was shown in Fig.2 and slope, intercept correlation factor and Regression equation were shown in Table.1

 

Fig.2: Linearity graph

 

Table1:LinearityData

Irbesartan Concentration (ppm)

Irbesartan Area

5.04ppm

0.1699

10.01ppm

0.3315

15.04ppm

0.5014

20.17ppm

0.6655

25.02ppm

0.8387

30.08ppm

0.9966

Slope

33.1022

Intercept

0.0019

Correlation factor

0.9999

 

Precision:-

The precision of  an  analytical method  is the degree of agreement among  individual test results when   the method is  applied repeatedly to multiple samplings of a homogeneous sample. The precision of an   analytical  method  is  usually   expressed  as   the standard deviation or relative standard deviation (coefficient of variation) of a series of measurements. Precision may  be  a  measure of either the degree of reproducibility or repeatability of the analytical method under normal operating conditions. The  ICH documents recommend that  repeatability should be assessed using  a minimum of nine  determinations covering the specified range for the procedure  (i.e., three concentrations and  three replicates of each concentration, or a minimum of six determinations at 100% of the test concentration).

 

System Repeatability:-

Standard solution is prepared at three difference concentration form different stock solution 15ppm,  20ppm and  25ppm were analysed and  taken absorbance on  triplicate and  RSD of absorbance at each  level  were calculated. The percentage RSD of  absorbance was  less  than 1.0%. Result are presented in Table.2

 

Table2:System Repeatability data

System Repeatability

SNo.

Concentration (ppm)

Absorbance

1

 

15.04

0.5032

2

0.5066

3

0.5029

 

RSD (%)

0.4

1

 

20.17

0.6737

2

0.6756

3

0.6737

 

RSD (%)

0.2

1

 

25.02

0.8387

2

0.8349

3

0.8276

 

RSD(%)

0.7

 

Method repeatability:-

Six preparation of Irbesartan sample was analyzed from sample preparation to final results by the same  analyst and the percentage RSD of obtained results was less than 2% and obtained result were within given range 100  ± 2. Result are presented in Table.2

 

System Reproducibility:- 

Three Irbesartan sample are analysed by this method in duplicate preparation and obtain result are in Table.4

 

Accuracy:-

It is defined as the closeness of test results obtained by the method to the true value. It may often be  expressed  as  percent  recovery by  the assay   of known,   added  amounts  of  analyte.  Accuracy  is  a measure of the exactness of the analytical method. The ICH documents recommend that accuracy be assessed using  a minimum of nine  determinations over a minimum of three concentration levels,  covering the specified range (i.e.,  three concentrations and  three replicates of each  concentration). The  three different concentrations of Irbesartan standard solutions were determined  from three  replicate analysis, using  the linear regression lines  (linearity section).  The deviations of the obtained results (expressed as percentage accuracy) were calculated from the true values  were presented in Table.5 The average deviations from true value are less than 2.0 %.

 

Range:-

The range obtained from Linearity, Precision and   Accuracy  is  summarized  -  ibesartan-5 ppm  to 30ppm  (25%  to  150%  of  nominal  sample concentration).

 

CONCLUSION

The method validation demonstrated that The Method is selective, precise, linear, and accurate for over the required concentration ranges of 5 to 30 ppm  (25 to150 % of Irbesartan nominal sample concentration) for quantitative analysis of irbesartan by UV spectroscopy.  So this method should be   used   in routine for quantitative analysis of irbesartan.

 

REFERENCE

1.        The  Merck  Index,  An  Encyclopaedia   of  Chemicals, Drugs and Biological, Thirteenth    Edition, Merck & Co. Inc., 2001, 13:3453.

2.        P  Palatini.  The  Journal  of  Clinical  Hypertension., 2005, 7, 96–101.

3.        USP   32:     "Validation    of    Compendial Methods".

4.        Guidance for Industry – Analytical Procedures and Method Validation (August-2000).

5.        W.D.  Snyder,  L.  Blumberg, in:  P.  Sandra,  M.L.  Lee (Eds.), Proceedings of the 14th International Symposium   on   Capillary   Chromatography, Baltimore, MD, May 1991, p. 28

6.        WHO,   Guidelines   for  Stability   Testing   of Pharmaceutical   Products    Containing    Well Established  Drug Substances  in Conventional Dosage Forms, in WHO Expert Committee

7.        Willard Hobart. H., Merritt L.L., Dean John. A., Instrumental Methods  of Analysis, 7th  edition, CBS Publishers, 580-610.

8.        International Conference on  Harmonization of Technical Requirement for Registration of Pharmaceuticals  for  Human  use,  ICH harmonized tripartite Guideline, Validation of Analytical procedures Text and methodology  Q2 (R1), 2005.

9.        R.  Jenke,  Chromatographic  Method  Validation:  A review   of  Current   Practices   and   Procedures.  I. General Concepts and Guidelines”, J. Liq. Chrom. and Rel. Technol., vol. 19 (1996), pp. 719-736.

10.     Albero,  I.,   Rodenas,  V.,   Garcia,  S.,   and   Sanchez, Pedreno, C., “Determination of Irbesartan in the presence of Hydrochlorothiazide by derivative Spectrophotometry”, J. Pharm. Biomed. Anal., 2004, 29 (1-2), pp. 229-305.

11.     Vetuchi, C., Giannandrea, A. Carlucci, G., and Mazzeo, P.”Determination of hydrochlorothiazide and Irbesartan in  Pharmaceuticals  by  fourth-order  UV derivative     Spectrophotometry,”    Farmaco,    2005, 60(8), pp.665-670.

12.     Gagigal  E,  Gonzalez  L,   Alonso  RM,  Jimenez   RM. Experimental design  methodologies to optimise  the spectrofluorimetric determination  of Losartan  and Valsartan in human urine. Talanta 2001;54:1121-33.

13.     Wang L, Asgharnejad M. Second derivative UV spectrometric determination of simvastatin in its tablet     dosage     form.    J     Pharm     Biomed    Anal 2000;21:1243-1248.

14.     Surekha  A,  Jain NK.  Difference  spectrophotometric estimation  of  amlodipine   besylate.   Indian   drugs 2000;37(7):351-353.

 

 

Received on 17.09.2015          Accepted on 20.10.2015        

© Asian Pharma Press All Right Reserved

Asian J. Pharm. Ana. 5(3): July- Sept. 2015; Page 147-150

DOI: 10.5958/2231-5675.2015.00024.1